| Specification: | V I B R A T I O N L E V E L M E T E R & A N A L Y S E R
Standards ISO 8041:2005
Meter mode RMS, VDV, MTVV or MAX, Peak, Peak-Peak
Simultaneous measurement in up to four channels with independent set of filters and detector constants
Analyser (option) 1/1 octave real time analysis, 15 filters with centre frequencies from 1 Hz to 16 kHz (Type 1, IEC 61260)
1/3 octave real time analysis, 45 filters with centre frequencies from 0.8 Hz to 20 kHz (Type 1, IEC 61260)
FFT real time analysis up to 1920 lines with Hanning, Kaiser-Bessel or Flat Top window
FFT cross spectra measurements
RPM rotation speed measurements parallel to the vibration measurement (1 ÷ 99999)
and more…
Filters Wd, Wk, Wc, Wj, Wm, Wb, Wg (ISO 2631), Wh (ISO 5349), HP1, HP3, HP10, Vel1, Vel3, Vel10,
VelMF, Dil1, Dil3, Dil10, KB
RMS & RMQ Detectors Digital True RMS & RMQ detectors with Peak detection, resolution 0.1 dB
Time Constants: from 100 ms to 10 s
Accelerometer (option) SV 39A/L seat accelerometer (100 mV/g sensitivity) for Whole-Body measurements
SV 50 set for Hand-Arm measurements (Dytran 3023M2 accelerometer, 10 mV/g sensitivity)
Other IEPE accelerometers
Measurement Range Accelerometer dependent, SV 39A/L seat accelerometer: 0.003 ms-2 RMS ÷ 500 ms-2 PEAK
Frequency Range 0.5 Hz ÷ 20 kHz; accelerometer dependent, with Dytran 3023M2: 2 Hz ÷ 10 kHz
S O U N D L E V E L M E T E R & A N A L Y S E R
Standards Type 1: IEC 61672-1:2002
Meter mode SPL, Leq, SEL, Lden, Ltm3, Ltm5, Statistics - Ln (L1-L99), LMax, LMin, LPeak
Simultaneous measurement in three profiles per channel (up to four channels)
with independent set of filters and detector time constants
Analyser (option) 1/1 octave real time analysis, 15 filters with centre frequencies from 1 Hz to 16 kHz (Type 1, IEC 61260)
1/3 octave real time analysis, 45 filters with centre frequencies from 0.8 Hz to 20 kHz (Type 1, IEC 61260)
Reverberation Time analysis in 1/3 octave bands (RT 60)
FFT real time analysis up to 1920 lines with Hanning, Kaiser-Bessel or Flat Top window
FFT cross spectra measurements
Sound Intensity Measurements
and more…
Weighting Filters Standard: A, C, Z and G
RMS Detector Digital True RMS detector with Peak detection, resolution 0.1 dB
Time Constants: Slow, Fast, Impulse
Microphone (option) SV 22, Type 1, 50 mV/Pa, prepolarised 1/2" condenser microphone with SV 12L preamplifier
SV 25, Type 2, dose meter, ceramic 1/2" microphone with integrated preamplifier
Measurement Range Total Dynamic Range: 17 dBA RMS ÷ 140 dBA Peak
Linearity Range: 24 dBA RMS ÷ 140 dBA Peak
Frequency Range 0.5 Hz ÷ 20 kHz; microphone dependent, with SV 22 microphone: 10 Hz ÷ 20 kHz
B A S I C D A T A
Input IEPE type (channels 1, 2, 3 - LEMO 4 pin & channel 4 - TNC connector)
Dynamic Range 100 dB, 4 x 20 bits A/D converters
Frequency Range 0.5 Hz ÷ 20 kHz, sampling rate 48 kHz
Data Logger Time History logging to internal memory or USB Memory Stick
Time domain signal recording on USB Memory Stick (option)
Display LCD 128 x 64 pixels plus icons with backlighting
Memory 32 MB non-volatile flash type, external USB Memory Stick (not included)
Interfaces USB 1.1 Client, USB 1.1 Host, RS 232 (option: SV 55 required), IrDA (option)
External I/O - AC output (1 V Peak) or Digital Input/Output (Trigger / Pulse)
Power Supply Four AA batteries (alkaline) operation time > 10 h (6.0 V / 1.6 Ah)
Four AA rechargeable batteries (not included) operation time > 14 h (4.8 V / 2.6 Ah)
SA 17A external battery pack (option) operation time > 24 h
External power supply 6 V DC ÷ 15 V DC (1.5 W)
USB interface 500 mA HUB
Environmental Conditions Temperature from -10 oC to 50 oC
Humidity up to 90 % RH, non-condensed
Dimensions 140 x 82 x 42 mm
Weight 510 grams with batteries
each function parallel to the meter mode with USB 1.1 Host function not active and backlight of |